Acquisition system
|
Number of input channels
|
4 analog
|
CH1, CH2, CH3, CH4
|
Isolated female BNC
|
Type
|
Differential input
|
Resolution
|
8, 12, 14, 16 bit user selectable
|
DC Accuracy
|
0.25 % (0.1 % typical) of full scale ± 1 LSB at 20°C to 25°C
To achieve rated accuracy, allow the
instrument to settle for 20 minutes. When subjected to extreme
temperatures, allow additional time for internal temperatures to
stabilize.
|
Ranges (full scale)
|
±200 mV
±400 mV
±800 mV
|
±2 V
±4 V
±8 V
|
±20 V
±40 V
±80 V
|
Coupling
|
AC/DC
|
Impedance
|
2 MOhm / 12 pF ± 1 %
1 MOhm / 20 pF ± 1 % when SafeGround enabled
|
Maximum voltage
|
200 V (DC + AC peak < 10 kHz)
|
Maximum Common Mode voltage
|
200 mV to 800 mV ranges : 2 V
2 V to 8 V ranges : 20 V
20 V to 80 V ranges : 200 V
|
Common Mode Rejection Ratio
|
-47 dB
|
Bandwidth
|
HS6D-1000
|
HS6D-500
|
HS6D-200
|
HS6D-100
|
HS6D-50
|
-3dB at 75 % of full scale input
|
250 MHz
|
250 MHz
|
250 MHz
|
100 MHz
|
100 MHz
|
Limit, selectable per channel
|
Off (250 MHz)
150 MHz
100 MHz
50 MHz
|
Off (250 MHz)
150 MHz
100 MHz
50 MHz
|
Off (250 MHz)
150 MHz
100 MHz
50 MHz
|
Off (100 MHz)
75 MHz
50 MHz
25 MHz
|
Off (100 MHz)
75 MHz
50 MHz
25 MHz
|
AC coupling cut off frequency (-3dB)
|
±1.5 Hz
|
SureConnect
|
Optionally available (option S)
|
Maximum voltage on connection
|
200 V (DC + AC peak < 10 kHz)
|
Resistance measurement
|
Optionally available (option S)
|
Ranges (Full scale)
|
100 Ω
200 Ω
500 Ω
|
1 kΩ
2 kΩ
5 kΩ
|
10 kΩ
20 kΩ
50 kΩ
|
100 kΩ
200 kΩ
500 kΩ
|
1 MΩ
2 MΩ
|
Accuracy
|
1 % of full scale
|
Response time (to 95%)
|
< 10 µs
|
SafeGround
|
Optionally available (option G)
|
Maximum voltage on connection
|
200 V (DC + AC peak < 10 kHz)
|
Maximum switch off current
|
500 mA
|
Response time
|
< 100 ns
|
|
Maximum sampling rate
|
HS6D-1000
|
HS6D-500
|
HS6D-200
|
HS6D-100
|
HS6D-50
|
8 bit
|
|
measuring one channel
|
1 GSa/s
|
500 MSa/s
|
200 MSa/s
|
100 MSa/s
|
50 MSa/s
|
measuring two channels
|
500 MSa/s
|
200 MSa/s
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
measuring three or four channels
|
200 MSa/s
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
10 MSa/s
|
12 bit
|
|
measuring one channel
|
500 MSa/s
|
200 MSa/s
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
measuring two channels
|
200 MSa/s
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
10 MSa/s
|
measuring three or four channels
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
10 MSa/s
|
5 MSa/s
|
14 bit
|
100 MSa/s
|
50 MSa/s
|
20 MSa/s
|
10 MSa/s
|
5 MSa/s
|
16 bit
|
6.25 MSa/s
|
3.125 MSa/s
|
1.25 MSa/s
|
625 kSa/s
|
312.5 kSa/s
|
|
Maximum streaming rate 1
|
HS6D-1000
|
HS6D-500
|
HS6D-200
|
HS6D-100
|
HS6D-50
|
8 bit
|
|
measuring one channel
|
200 MSa/s2
|
100 MSa/s2
|
40 MSa/s
|
20 MSa/s
|
10 MSa/s
|
measuring two channels
|
100 MSa/s3
|
50 MSa/s3
|
20 MSa/s
|
10 MSa/s
|
5 MSa/s
|
measuring three or four channels
|
50 MSa/s4
|
25 MSa/s4
|
10 MSa/s
|
5 MSa/s
|
2.5 MSa/s
|
12 bit
|
|
measuring one channel
|
100 MSa/s3
|
50 MSa/s3
|
20 MSa/s
|
10 MSa/s
|
5 MSa/s
|
measuring two channels
|
50 MSa/s4
|
25 MSa/s4
|
10 MSa/s
|
5 MSa/s
|
2.5 MSa/s
|
measuring three or four channels
|
25 MSa/s5
|
12.5 MSa/s5
|
5 MSa/s
|
2.5 MSa/s
|
1.25 MSa/s
|
14 bit
|
|
measuring one channel
|
100 MSa/s3
|
50 MSa/s3
|
20 MSa/s
|
10 MSa/s
|
5 MSa/s
|
measuring two channels
|
50 MSa/s4
|
25 MSa/s4
|
10 MSa/s
|
5 MSa/s
|
2.5 MSa/s
|
measuring three or four channels
|
25 MSa/s5
|
12.5 MSa/s5
|
5 MSa/s
|
2.5 MSa/s
|
1.25 MSa/s
|
16 bit
|
6.25 MSa/s6
|
3.125 MSa/s
|
1.25 MSa/s
|
625 kSa/s
|
312.5 kSa/s
|
1.
2. ≤ 40 MSa/swhen connected to USB 2.0
3. ≤ 20 MSa/swhen connected to USB 2.0
4. ≤ 10 MSa/swhen connected to USB 2.0
5. ≤ 5 MSa/swhen connected to USB 2.0
6. ≤ 3.125 MSa/swhen connected to USB 2.0
|
Sampling clock source
|
|
Internal
|
TCXO
|
Accuracy
|
± 0.0001 %
|
Stability
|
± 1 ppm over 0°C to 55°C
|
Time base aging
|
±1 ppm/year
|
External
|
LVDS, on CMI connectors
|
Input frequency
|
10 MHz ± 1 %
16.369 MHz ± 1 %
|
|
Memory
|
Standard model
|
XM option
|
8 bit resolution
|
|
Measuring one channel
|
1 Mpts per channel
|
256 Mpts per channel
|
Measuring two channels
|
512 Kpts per channel
|
128 Mpts per channel
|
Measuring three or four channels
|
256 Kpts per channel
|
64 Mpts per channel
|
12, 14, 16 bit resolution
|
|
Measuring one channel
|
512 Kpts per channel
|
128 Mpts per channel
|
Measuring two channels
|
256 Kpts per channel
|
64 Mpts per channel
|
Measuring three or four channels
|
128 Kpts per channel
|
32 Mpts per channel
|
|
Trigger
|
|
System
|
Digital, 2 levels
|
Source
|
CH1, CH2, CH3, CH4, OR, digital external
|
Trigger modes
|
Rising edge, falling edge, any edge, inside window, outside window, enter window, exit window, pulse width
|
Level adjustment
|
0 to 100 % of full scale
|
Hysteresis adjustment
|
0 to 100 % of full scale
|
Resolution
|
0.006 % (14, 16 bits) / 0.025% (12 bits)
|
Pre trigger
|
0 to selected record length, 1 sample resolution
|
Post trigger
|
0 to selected record length, 1 sample resolution
|
Trigger hold-off
|
0 to 63 MSamples, 1 sample resolution
|
Trigger delay
|
0 to 8 GSamples, 1 sample resolution
|
Digital external trigger
|
|
Input
|
Extension connector pins 1 and 2
|
Range
|
0 to 2.5 V (TTL)
|
Coupling
|
DC
|
Jitter
|
≤ 1 sample
|
Multi-instrument synchronization
|
Maximum number of instruments
|
Limited by number of available USB ports
|
Synchronization accuracy
|
0 ppm
|
CMI interface
|
2x, CMI 1, CMI 2
|
Required coupling cable
|
TP-C50H Coupling cable CMI
|
Maximum coupling cable length
|
50 cm
|
Probe calibration
|
Output
|
Extension connector pins 3 (signal) and 6 (ground)
|
Signal
|
Square wave
|
Level
|
-1 V to 1 V
|
Frequency
|
1 kHz
|
Interface
|
Interface
|
USB 3.0 SuperSpeed (5 Gbit/s);
(USB 2.0 HighSpeed compatible)
|
Network support
|
Yes, via TPISS Instrument Sharing Server
|
Power Requirements
|
Power
|
From USB port or external input
|
Consumption
|
5 VDC 1200 mA max
|
External power
|
From second USB port or power adapter
|
Physical
|
Instrument
|
|
Height
|
25 mm (1 inch)
|
Length
|
170 mm (6.7 inch)
|
Width
|
140 mm (5.2 inch)
|
Weight
|
500 g (17.6 ounce)
|
USB cord length
|
1.8 m (70 inch)
|
I/O connectors
|
Channel 1, 2, 3, 4
|
Isolated BNC
|
USB
|
Fixed cable with USB 3.0 type A connector, 1.8 m
|
Extension connector
|
D-sub 9 pins female
|
Power
|
3.5 mm power socket
|
CMI I/O connectors
|
2 x HDMI type C socket
|
System requirements
|
PC I/O connection
|
USB 2.0, USB 3.0 or USB 3.1
|
Operating system
|
Windows 7/8/10, 32 and 64 bits
Linux (via LibTiePie SDK)
|
Environmental conditions
|
Operating
|
|
Ambient temperature
|
20°C to 25°C (10°C to 40°C without specifications)
|
Relative humidity
|
10 % to 90 %, non condensing
|
Storage
|
|
Ambient temperature
|
-20°C to 70°C
|
Relative humidity
|
5 % to 95 %, non condensing
|
Certifications and Compliances
|
CE mark compliance
|
Yes
|
RoHS
|
Yes
|
EN 55011:2009/A1:2010
|
Yes
|
EN 55022:2006/A1:2007
|
Yes
|
EN 61000-6-1:2007
|
Yes
|
EN 61000-6-3:2007
|
Yes
|
Package contents
|
Instrument
|
Handyscope HS6 DIFF
|
Measure leads
|
4 x Measure lead TP-C812B, Differential BNC -> 4 mm banana plug
|
Differential attenuator
|
4 x Differential attenuator TP-DA10
|
Alligator clips, large
|
Alligator Clip TP-AC80I - Set, 8 large alligator clips
|
Alligator clips, medium
|
Alligator Clip TP-AC10I - Set, 8 medium alligator clips
|
Alligator clips, small
|
Alligator Clip TP-AC5 - Set, 8 small alligator clips
|
Accessories
|
Handyscope power supply, external power cable for USB port, Probe compensation cable TP-DB9-BNC-30 (only with option SafeGround)
|
Software
|
for Windows 7/8/10, via website
|
Drivers
|
for Windows 7/8/10, via website
|
Manuals
|
instrument manual and software user's manuals
color printed and digital, via website
|
Optional Accessories
|
Optional accessories for the Handyscope HS6 DIFF need to be ordered separately
|
Oscilloscope probe
|
Oscilloscope Probe 1:1-1:10 - HP-3250I, only with SafeGround option G
|
Measure lead single ended
|
Measure lead TP-BNCI-100, only with SafeGround option G
|
Customer service
|
TiePie engineering instruments are designed,
manufactured and tested to provide high reliability. In the unlikely
event you experience difficulties, the TiePie engineering instruments
are fully warranted for two years.
|
This warranty includes
|
All parts and labor (excluding probes and/or measure leads)
No charge for return shipping
Long-term 7 year support
Upgrade to the latest software at no charge
|
Optionally available
|
Extension to five year warranty, option W5
|